Selection of internal standards for analysis of silicate rocks and limestones by laser ablation inductively coupled plasma atomic emission spectrometry

Logo poskytovatele

Varování

Publikace nespadá pod Fakultu sportovních studií, ale pod Přírodovědeckou fakultu. Oficiální stránka publikace je na webu muni.cz.
Autoři

KANICKÝ Viktor MERMET Jean - Michel

Rok publikování 1996
Druh Článek v odborném periodiku
Časopis / Zdroj Fresenius Journal of Analytical Chemistry
Fakulta / Pracoviště MU

Přírodovědecká fakulta

Citace
Obor Analytická chemie, separace
Klíčová slova Inductively Coupled Plasma; Atomic Emission Spectrometry; Laser Ablation; Silicates; Lithiumtetraborate Fusion;
Popis Included (Si, Ca) and added (Li, B) internal standards (IS) have been used comparatively in LA-ICP AES of silicates and limestones to improve both precision and accuracy. A Q-switched Nd:YAG laser (355 nm, 10 Hz, 10 mJ per shot) has been applied for ablation, Samples have been prepared by fusion with Li2B4O7 and measured using a Perkin-Elmer Optima 3000 ICP system, Both types of IS have given a relative standard deviation (RSD) less than 2-1%, which improved the repeatability by a factor of 2-10 and calibration graphs have been linear over the whole concentration range.
Související projekty:

Používáte starou verzi internetového prohlížeče. Doporučujeme aktualizovat Váš prohlížeč na nejnovější verzi.

Další info