Amperometric sensor for detection of chloride ions

Warning

This publication doesn't include Faculty of Sports Studies. It includes Faculty of Science. Official publication website can be found on muni.cz.
Authors

TRNKOVÁ Libuše ADAM Vojtěch HUBÁLEK Jaromír BABULA Petr KIZEK René

Year of publication 2008
Type Article in Periodical
Magazine / Source Sensors
MU Faculty or unit

Faculty of Science

Citation
Field Electrochemistry
Keywords chloride ions; silver; voltammetry; amperometry
Description Chloride ion sensing is important in many fields such as clinical diagnosis, environmental monitoring and industrial applications. We have measured chloride ions at a carbon paste electrode (CPE) and at a CPE modified with solid AgNO3, a solution of AgNO3 and/or solid silver particles. Detection limits (3 S/N) for chloride ions were 100 mu M, 100 mu M and 10 mu M for solid AgNO3, solution of AgNO3 and/or solid silver particles, respectively. The CPE modified with silver particles is the most sensitive to the presence chloride ions. After that we approached to the miniaturization of the whole electrochemical instrument. Measurements were carried out on miniaturized instrument consisting of a potentiostat with dimensions 35 x 166 x 125 mm, screen printed electrodes, a peristaltic pump and a PC with control software. Under the most suitable experimental conditions (Britton-Robinson buffer, pH 1.8 and working electrode potential 550 mV) we estimated the limit of detection (3 S/N) as 500 nM.
Related projects:

You are running an old browser version. We recommend updating your browser to its latest version.

More info