Comment on “Hypersensitive Luminiscence of Eu3+ in Dimethyl Sulfoxide As a New Probing for Water Measurement”

Investor logo

Warning

This publication doesn't include Faculty of Sports Studies. It includes Central European Institute of Technology. Official publication website can be found on muni.cz.
Authors

PEKÁRKOVÁ Pavla LUBAL Přemysl TÁBORSKÝ Petr

Year of publication 2012
Type Article in Periodical
Magazine / Source Analytical Chemistry
MU Faculty or unit

Central European Institute of Technology

Citation
Web pdf
Doi http://dx.doi.org/10.1021/ac301934d
Field Analytic chemistry
Keywords europium(III) luminiscence; luminiscence decay; water determination; organic solvent; ionic liquid
Description Yao and Chen(see Analytical Chemistry (Washington, DC, United States) (2011), 83 (6), 1879-1882) have shown a method for determination of water traces in dimethyl sulfoxide (DMSO). The described method is interesting, but in our opinion, it can be entirely replaced by other methods using the same reagents and based on measurement of Eu(III) luminescence lifetimes. Moreover, there is quite a lot of inaccuracy in the article which we would like to comment. Finally, we want to briefly show a few other possible applications of the method described in this comment.